Service
DC Gain Accuracy Test
4-8
Figure 4-1
4 Select the Default Setup item in the Save/Recall menu.
5 Press the
Acquire front panel button.
6 Select the mode item in the Acquire menu until Average appears.
7 Select the
Averages item in the Acquire menu until 256 appears.