EasyManuals Logo
Home>Agilent Technologies>Test Equipment>3000 Series

Agilent Technologies 3000 Series User Manual

Agilent Technologies 3000 Series
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #71 background imageLoading...
Page #71 background image
Utility Controls
Self-Calibration
2-53
Self-Calibration
Pressing the Self-Cal menu key starts the automatic calibration routine which
adjusts the internal circuitry of the oscilloscope for the best measurement
accuracy. The the automatic calibration should be run when the ambient
temperature changes by 5 ºC or more.
Figure 2-29
Calibration Dialog Box
Before performing the automatic calibration, allow the oscilloscope to warm-up at
least 30 minutes.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 3000 Series and is the answer not in the manual?

Agilent Technologies 3000 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals