3-2
Specifications and Characteristics
Specifications
Specifications
All specifications are warranted. Specifications are valid after a 30-minute warm-up
period and ±5 °C from last calibration temperature.
Bandwidth (-3dB) DSO3062A: 60 MHz
DSO3102A: 100 MHz
DSO3152A: 150 MHz
DSO3202A: 200 MHz
DC Vertical Gain Accuracy 2 mV/div to 5 mV/d: ±4.0% full scale
10 mV/div to 5 V/div: ±3.0% full scale