Service
Analog Bandwidth - Maximum Frequency Check
4-15
Figure 4-5
5 Select the Default Setup item in the Save/Recall menu.
6 Press the
Autoscale front panel button.
7 Set the channel 1 vertical scale to 200 mV/div.
Figure 4-6
8 Set the horizontal scale to 500 ns/div.