EasyManuals Logo
Home>Agilent Technologies>Test Equipment>54831A

Agilent Technologies 54831A Service Guide

Agilent Technologies 54831A
170 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #106 background imageLoading...
Page #106 background image
Chapter 5: Troubleshooting
To Configure the motherboard jumpers and setup BIOS
5–21
BIO
S Setup Procedure
1
Go to Load Setting 2 Defaults and press Enter key. Select Y to load the defaults of
BIOS Setting 2 and press the Enter key.
2 If you have a power board part number 54855-66502 then perform the following steps.
Otherwise, go to step 7.
3 Go to Power Management Setup and press the Enter key.
4 Select the PWORN After PWR-Fail setting and press the Enter key.
5 Select the Off option and press the Enter key.
6 Return to the CMOS Setup Utility by pressing the Esc key.
7 Press F10 to save and exit the setup. Type “Y” to save changes.

Table of Contents

Other manuals for Agilent Technologies 54831A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 54831A and is the answer not in the manual?

Agilent Technologies 54831A Specifications

General IconGeneral
BrandAgilent Technologies
Model54831A
CategoryTest Equipment
LanguageEnglish

Related product manuals