Chapter 3: Testing Performance
To verify threshold accuracy
3–32
Digit
al
Channels
Threshold
Accuracy
(54830/31/32/33
D)
±100 mV +3% of threshold setting Pass/Fail
Specification Limits Channel D7-D0 Channel D15-D8
5 V -250 mV 4.750 V __________________ __________________
5 V +250 mV 5.250 V __________________ __________________
-5 V -250 mV -5.250 V __________________ __________________
-5 V +250 mV -4.750 V __________________ __________________
0 V -100 mV -100 mV __________________ __________________
0 V +100 mV 100 mV __________________ __________________
Test Limits Results