EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>6420

Agilent Technologies 6420 User Manual

Agilent Technologies 6420
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #43 background imageLoading...
Page #43 background image
Agilent Triple Quadrupole MS and Sensitivity 3
Example of chemical noise reduction
Agilent 6400 Series Triple Quad LC/MS Concepts Guide 43
Example of chemical noise reduction
The Agilent 6400 Series Triple Quadrupole mass spectrometer
passes through four transitional steps in translating a signal in
the MRM process (Figure 18).
Step 1 The spectrum at the far left represents everything that
is being ionized at the ion source. This example shows the ESI
spectrum of a phenylurea pesticide. A triple quadrupole mass
spectrometer reduces chemical noise for low-level quantitation
in a dirty matrix more than a single quadrupole LC/MS does.
Step 2 This step is accomplished by first selecting the pesticide
of interest at m/z 210 from the co-eluting interferences seen in
the rest of the spectrum. The second spectrum shows the result
after passing through the first quadrupole, or MS1 (Q1).
Figure 18 Multiple reaction monitoring (MRM)

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6420 and is the answer not in the manual?

Agilent Technologies 6420 Specifications

General IconGeneral
BrandAgilent Technologies
Model6420
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals