49
A
Specifications
Specifications
Table A-1 lists the specifications of the dc source. Specifications are warranted over the ambient
temperature range of 0 to 55 °C. Unless otherwise noted, specifications apply when measured at the rear
terminals after a 30-minute warm-up period.
Table A-1. Performance Specifications
Parameter Agilent
6631B
Agilent 6632B
Agilent
66332A
Agilent
6633B
Agilent
6634B
Output Ratings
Voltage:
Current:
0 8V
0 10A
0 20 V
0 5A
0 50 V
0 2A
0 100 V
0 1A
Programming Accuracy
(@ 25C ±5°C)
Voltage:
+Current:
0.05% + 5mV
0.05% + 4mA
1
0.05% + 10mV
0.05% + 2mA
1
0.05% + 20mV
0.05% + 1mA
1
0.05% + 50mV
0.05% + 0.5
1
mA
DC Measurement
Accuracy
(via GPIB or front panel
meters with respect to
actual output at 25C
±5°C)
Voltage:
Low Current range
20 mA to +20 mA:
High Current range
+20mAto+ratedI:
20 mA to rated I:
0.03% + 2 mV
0.1% + 2.5 µA
2
0.2% + 1 mA
3
0.2% + 1.6 mA
0.03% + 3 mV
0.1% + 2.5 µA
2
0.2% + 0.5 mA
3
0.2% + 1.1 mA
0.03% + 6 mV
0.1% + 2.5 µA
2
0.2% + 0.25 mA
3
0.2% + 0.85 mA
0.03% + 12 mV
0.1% + 2.5 µA
2
0.2% + 0.25 mA
3
0.2% + 0.85 mA
Ripple and Noise
(in the range of 20 Hz to
20 MHz with outputs
ungrounded or with
either terminal grounded)
Normal mode voltage
(rms/p-p):
Fast mode voltage
(rms/p-p):
Current (rms):
0.3mV/3mV
3
1mV/10mV
3mA
0.3mV/3mV
4
1mV/10mV
2mA
0.5mV/3mV
4
1mV/15mV
2mA
0.6mV/5mV
4
2mV/25mV
2mA
Load Regulation
5
(change in output voltage
or current for any load
change within ratings)
Voltage:
Current:
2mV
2mA
2mV
1mA
4mV
1mA
5mV
1mA
Line Regulation
(change in output voltage
or current for any line
change within ratings)
Voltage:
Current:
0.5 mV
1mA
0.5 mV
0.5 mA
1mV
0.25 mA
1mV
0.25 mA
Transient Response
Time
5
Normal mode:
Fast mode:
< 100 µs
<50µs
(for the output voltage to recover to its previous level within 0.1% of the voltage rating of the unit or 20 mV following a change
in load current of up to 50% of the output current rating)
1
0.05% +6.63mA ('31B), 3.32mA ('32B/332A), 1.53mA ('33B), 0.76mA ('34B) when programming between zero and 0.03% of full scale current
2
This specification may degrade slightly when the unit is subjected to an RF field 3 V/meter.
3
For Agilent 66332A: applies in SCPI mode, with current detector set to DC. With current detector set to ACDC, accuracy is 0.2% + four times
the fixed error value. In COMPatibility mode, accuracy is 0.2% + six times the fixed error value.
4
For Agilent 6631B, 6632B, and 66332A (from 1 MHz to 20 MHz ) = 0.3mV/15mV;
For Agilent 6633B and 6634B (from 1 MHz to 20 MHz ) = 0.5mV/15mV
5
Applies at rear terminals with unit set to remote sensing and with sense terminals externally jumpered to their respective output terminals.
For units with Option 760, Normal mode transient response is as follows: < 250µs ('32B/332A); < 230µs ('33B/34B),

Table of Contents

Other manuals for Agilent Technologies 6633B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6633B and is the answer not in the manual?

Agilent Technologies 6633B Specifications

General IconGeneral
BrandAgilent Technologies
Model6633B
CategoryPower Supply
LanguageEnglish

Related product manuals