EasyManua.ls Logo

Agilent Technologies 7890A - NPD Cleanliness

Agilent Technologies 7890A
18 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Replacing the NPD Bead
Instruction Guide 17
NPD Cleanliness
Below is an overview of typical NPD leakage currents for a
system set to recommended flows and temperature.
When changing insulators and seals, cool the detector to 75 °C
or below. This will minimize the chance of stripping the three
T- 20 Torx screws.
The installation of traps is recommended for the detector,
hydrogen, air, and make- up gas. A moisture trap followed by
chemical traps is highly recommended.
pA Leakage Current System status Comment
0.3-0.9 CLEAN
1.0+ Dirty Decrease in sample response.
10+ Very dirty Almost no sample response.
Cause is usually contaminated ceramic
insulators on the collector - usually the
upper insulator and seal.
NPD Bead Replacement.book Page 17 Monday, October 13, 2008 2:32 PM

Other manuals for Agilent Technologies 7890A

Related product manuals