EasyManua.ls Logo

Agilent Technologies 8565EC

Agilent Technologies 8565EC
512 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
462 Chapter9
IF Section
A5 IF Assembly
Common IF Signature Problems
Region A of Figure 9-9, “Detailed IF Adjust Signature (1)” is
noisy:
Suspect the first LC pole.
Region B of Figure 9-9 on page 459 is flat:
Suspect the third step-gain stage, the fine attenuator, or the fourth
LC-pole output amplifier.
Region C of Figure 9-9, “Detailed IF Adjust Signature (1)” has
no 15 dB step:
Suspect the reference 15 dB attenuator.
Region D of Figure 9-9 on page 459 is flat:
Suspect the second step-gain stage.
Entire signature noisy:
If the signature resembles Figure 9-14 on page 463, suspect a broken
first step-gain stage or a break in the signal path in the input switch,
first crystal pole, or second crystal pole.
Correct shape but noisy:
If the signature resembles Figure 9-15, “Noise with Correct Shape”
suspect the second crystal-pole output amplifier.
Amplitude of Region B of Figure 9-16, “Region B Amplitude
Variation” varies more than 12 dB:
Suspect the third step-gain stage output amplifier.
Region B of Figure 9-17, “Region B Amplitude Offset” is kinked:
Suspect the fourth LC-pole output amplifier.

Table of Contents

Related product manuals