Service Guide 11-5
8719ET/20ET/22ET Error Terms
8719ES/20ES/22ES Error Term Inspection
RF Network Analyzers
Error Term Inspection
Table 11-1 (ES models) and Table 11-2 (ET models) list the calibration coefficients along
with their corresponding test numbers. Refer to the appropriate table when performing
this procedure.
Table 11-1 Calibration Coefficient Terms and Tests for ES Models
Calibration
Coefficient
Calibration Type Test Number
Response Response and
Isolation*
1-port 2-port
1E
R
or E
T
E
X
(E
D
)E
D
E
DF
31
2E
T
(E
R
)E
S
E
SF
32
3E
R
E
RF
33
4E
XF
34
5E
LF
35
6E
TF
36
7E
DR
37
8E
SR
38
9E
RR
39
10 E
XR
40
11 E
LR
41
12 E
TR
42
NOTES:
Meaning of first subscript: D=directivity; S=source match; R=reflection tracking;
X=crosstalk; L=load match; T=transmission tracking.
Meaning of second subscript: F=forward; R=reverse.
* Response and Isolation cal yields: E
X
or E
T
if a transmission parameter (S
21
, S
12
) or E
D
or
E
R
if a reflection parameter (S
11
, S
22
)