12-12 Service Guide
Theory of Operation 8719ET/20ET/22ET
Digital Control Theory 8719ES/20ES/22ES
RF Network Analyzers
TEST SEQ. This outputs a TTL signal that can be programmed in a test sequence to be
high or low, or pulse (10 µ seconds) high or low at the end of a sweep for a robotic part
handler interface.
LIMIT TEST. This outputs a TTL signal of the limit test results as follows:
Pass: TTL high
Fail: TTL low
VGA OUTPUT. This provides a video output of the analyzer display that is capable of
being viewed on VGA monitor.