2-40 Service Guide
System Verification and Performance Tests 8719ET/20ET/22ET
Performance Tests 8719ES/20ES/22ES
RF Network Analyzers
8722ET/ES Models
1. Press .
2. Press .
3. Refer to Table 2-10 and set the analyzer to the power value that is listed in the P
Ref
row
for the particular analyzer under test. After you enter the value, press .
4. Wait for analyzer to complete one full sweep.
5. Press .
6. Press
.
Recording P
1
Data
7. Press .
8. Enter the power value that is listed in the P
1
row of Table 2-10 for the particular
analyzer under test. After you enter the value, press .
9. Wait for analyzer to complete one full sweep.
10.Press .
11.Read the marker value from the analyzer display and add 5 dB. Record the calculated
value in the 8722ET/ES (0.05 20 GHz) “P
1
Max.” column of the “Source Linearity
performance test record.
12.Press .
13.Read the marker value from the analyzer display and add 5 dB. Record the calculated
value in the 8722ET/ES (0.05 20 GHz) “P
1
Min.” column of the “Source Linearity”
performance test record.
Recording P
1
Data for the 20 GHz to 40 GHz Band
14.To record data for P
1
data in the 20 GHz to 40 GHz band, press .
15.Press .
16.Repeat steps 2 through 13 of the section, except record the values in the 8722ET/ES
(20 40 GHz)P
1
Max.” and “P
1
Min.” columns of the “Source Linearity” performance
test record.
Stop 20 G/n
Sweep Setup
POWER
x1
Display
DATA MEMORY
DATA/MEM
Marker
MODE MENU
FIXED MKR POSITION
FIXED MKR VALUE
0 x1
Sweep Setup
POWER
x1
Marker Search
SEARCH: MAX
SEARCH: MIN
Start 20 G/n
Stop 40 G/n