7-56
Operating Concepts
Calibration Routines
E-CAL
The E-Cal calibration menu is activated by pressing in the calibration
menu.
The E-Cal (Electronic Calibration) system determines systemic errors of the analyzer
through a one-time connection of an E-Cal module to the network analyzer ports. The
random error of connector repeatability is reduced substantially through a one-time
connection when compared to frequent connections and disconnections of the conventional
short/open/load methods.
E-CAL MENU