4-8
Supplement for the User’s Guide (Firmware Version 7.68)
Procedures for Error Correcting Your Measurements
Procedures for Error Correcting Your Measurements
Types of Error Correction
Purpose and Use of Different Error Correction Procedures
Correction Procedure Corresponding
Measurement
Errors Corrected Standard Devices
Enhanced Response
and Enhanced
Reflection
Transmission or
reflection
measurement when
improved accuracy is
desired. Not as
accurate as 2-port
calibration.
Directivity, source
match, and frequency
response for
reflection. Frequency
response, source
match, and isolation
for transmission.
Enhanced reflection
corrects for load
match on bilateral
devices.
Short, open, load, and
thru or ECal module.