Introduction
Agilent 240/280 Series AA User’s Guide 31
Specifications
Your Agilent AA instrument is designed for indoor use. It is suitable
for the following categories:
Installation category II
Pollution degree 2
Equipment Class I
Environmental
Condition Altitude Temp
t (°C)
Humidity (%RH)
non-condensing
Non-operating (transport) 0-2133 m
(0-7000')
5-45 20-80
Operating within
performance specifications
0-853 m
(0-2800')
10-35 8-80
853-2133 m
(2800-7000')
10-25 8-80
For optimum analytical performance it is recommended that the
ambient temperature of the laboratory be between 20-25 °C and be
held constant to within ±2 °C throughout the entire working day.