Introduction
Agilent 240/280 Series AA User’s Guide 33
Power Connections
Agilent AA instruments are supplied with different power plugs to
suit the area into which they are sold:
AA instrument GTA 120, Zeeman
Area - code Plug supplied
Australia -00 10 A, 250 VAC
Complies with AS3112
Clipsal 439D15M
USA -01 Complies with NEMA 5-15P Complies with NEMA L6-30P
(Hubbell #2621)
Canada -01 Complies with NEMA 5-15P 20 A, 250 VAC, Complies with
NEMA L6-20P
(Hubbell #2321+)
Europe -02 Perena 3410 (Complies with CEE 7
Sheet VII or NFC 61.303)
Kaiser CEBEC 616 VDE
(Complies with DIN 49441R2)
Suitable socket required
Australia -00 General purpose 10 A 250 V outlet
(HPM 787, Clipsal 15)
Dedicated circuit, 15 A 250 V
outlet
(HPM 787/15, Clipsal 15/15)
USA -01 To comply with NEMA 5-15R (15 A
supply) (Hubbell IG 5262)
To comply with NEMA L6-30R
(30 A supply)
(Hubbell #2626)
Canada -01 To comply with NEMA 5-15R (15 A
supply) (Hubbell IG 5262)
20 A, 250 VAC, Complies with
NEMA L6-20R
(Hubbell #2326+)
Europe -02 To comply with CEE 7 standard No.7
Sheet VII, or Norma Francais C61.303
Sheet V.A.
No standard known (Kaiser CEBEC
702 type 31/131.5)
Power supply, current rating
and overload protection
Between 5 and 20 A Between 30 and 40 A
Power supply Single phase Single phase