Installation
Agilent 240/280 Series AA User’s Guide 43
If you are running an Agilent AA-DUO system (simultaneous
flame and furnace operation):
You must now set up the second instrument. To do this:
a Select the DUO check box.
b Select the Instrument #2 tab.
c Select the instrument type from the drop-down list. It does
not matter which instrument is assigned to which page,
however, ‘Instrument #1’ must have the IEEE “9” setting and
‘Instrument #2’ must have the IEEE “8” setting. The IEEE
settings are configured after installing SpectrAA PRO. Take
note of which instrument is listed in ‘Instrument #2’. See the
following image.
11 If you are using any accessories, click the Accessories tab in the
‘SpectrAA Instrument Configuration’ window. Select your GTA
(Graphite Tube Atomizer), Flame Autosampler and /or ETC 60
from the drop-down lists.
a If you select GTA 120, the option for High capacity racks is
enabled. Select this option to enable support for the optional
high capacity carousel (130 x 1.1 mL samples and 5 x 10 mL
standard/modifier solutions).