Installation
48 Agilent 240/280 Series AA User’s Guide
5 From the ‘Choose Destination Location’ window, select the drive
or sub-directory where the SpectrAA software will be installed. If
you wish to install the file in a directory other than the default
location, select Browse.
6 From the Select Program Folder window choose the name the
program shortcut will use on the Windows Start menu.
7 On the Start Copying Files window, check that all settings are
correct. Select Next to begin the installation.
8 From the ‘SpectrAA Instrument Configuration’ window, select
your instrument type from the drop down list on the
‘Instrument #1’ page. Then do the following:
If you are running a one instrument system:
Continue with Step 9.
If you are running an Agilent AA-DUO system (simultaneous
flame and furnace operation):
You must now set up the second instrument. To do this:
a Select the DUO check box.
b Select the Instrument #2 tab.
c Select the instrument type from the drop-down list. It does
not matter which instrument is assigned to which page,
however, ‘Instrument #1’ must have the IEEE “9” setting and
‘Instrument #2’ must have the IEEE “8” setting. The IEEE
settings are configured after installing SpectrAA CFR. Take
note of which instrument is listed in ‘Instrument #2’. See the
following image.