Introduction
Agilent Cary 600 Series FTIR Spectrometers and Microscopes User’s Guide 25
Table 2. Cooling methods for available detectors
Detector Cooling method
High-sensitivity MCT Liquid nitrogen
Linearized high-sensitivity MCT Liquid nitrogen
Linearized broadband MCT Liquid nitrogen
InSb Liquid nitrogen
Room temperature DLaTGS None
Cooled DLaTGS Peltier
Far-infrared DLaTGS None
Si None
PbSe None
InGaAs None
Installation requirements
Before receiving your Agilent Cary 600 Series FTIR instrument, you
will have been given a Site Preparation Guide, which describes the
environmental and operating requirements of the Cary 600 Series
FTIR system. You must prepare your laboratory according to these
instructions before the Cary 600 Series FTIR can be installed. You
should keep the Site Preparation Guide for future reference. If you
have misplaced your copy, you can obtain a replacement from your
local Agilent office.
Training
The software provided with your Agilent Cary 600 Series FTIR
spectrometer and microscope includes extensive Help.
If Agilent installs the instrument(s), the Agilent representative will
demonstrate the basic operating procedures while conducting the
installation performance tests during the installation procedure. The
Agilent representative, however, is not necessarily experienced in
complex analytical routines and is not authorized to conduct
extensive training.