102 Chapter 2
Instrument Functions: A - L
Display
Instrument Functions: A - L
Remote Command:
:CALCulate:LLINe[1]|2:MARGin:STATe OFF|ON|0|1 turns on margins on or off. If the
margin and limit display are both turned off, limit test is automatically turned off.
:CALCulate:LLINe[1]|2:MARGin:STATe?
Responds with the margin state; 0 = off 1 = on.
:CALCulate:LLINe[1]|2:MARGin <ampl_rel>
Defines the amount of measurement margin that is added to the designated limit line.
:CALCulate:LLINe[1]|2:MARGin?
Responds with the margin offset value.
2.5.5.1.5 Edit
Pressing
Edit accesses menus for editing limit lines and for accessing the limit-line table
editor. Navigation through the limit-line table is achieved by using the front-panel arrow
and tab keys. Entering data in each field navigates to the next field. New limit segments
will only be applied after the editor is closed.
Pressing
Return, or any key that is not associated with the editor, will close the limit-line
table editor.
NOTE Refer to your Getting Started guide for more information about navigation
within tables using your front-panel arrow and tab keys.
Key Path:
Display, Limits, Limit 1
Display
, Limits, Limit 2
Factory Preset
and *RST: Exits the edit mode.
History: Added with firmware revision A.03.00

Table of Contents

Other manuals for Agilent Technologies E4446A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4446A and is the answer not in the manual?

Agilent Technologies E4446A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4446A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals