Chapter 6 367
Using the STATus System
Common Using the STATus System IEEE Commands
Using the STATus System
Recall
*RCL <register>
This command recalls the instrument state from the specified
instrument memory register.
If the state being loaded has a newer firmware revision than the
revision of the instrument, no state is recalled and an error is reported.
If the state being loaded has an equal firmware revision than the
revision of the instrument, the state will be loaded.
If the state being loaded has an older firmware revision than the
revision of the instrument, the instrument will only load the parts of
the state that apply to the older revision.
Key Type: There is no equivalent front-panel key.
Range: registers are an integer, 0 to 127
Remarks: See also commands :MMEMory:LOAD:STATe and
:MMEMory:STORe:STATe
Example: *RCL 12
Front Panel
Access:
File, Recall State
Reset
*RST
This command presets the instrument to a factory defined condition
that is appropriate for remote programming operation. In Spectrum
Analysis Mode *RST is equivalent to performing the commands
:SYSTem:PRESet, with preset type set to MODE.
*CLS which clears the STATus bits and error queue
*RST does not change the mode and only resets the parameters for the
current mode.
The :SYSTem:PRESet command is equivalent to a front panel
Preset
key.

Table of Contents

Other manuals for Agilent Technologies E4446A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4446A and is the answer not in the manual?

Agilent Technologies E4446A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4446A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals