284 Chapter 4
Instrument Functions: P - Z
Trig
Instrument Functions: P - Z
4.11.6 RF Burst (Wideband)
Allows the analyzer to be triggered by an RF burst envelope signal.
Key Path:
Trig
Remote Command:
See
Trig on page 281
Example: TRIG:SOUR RFB
4.11.7 Trig Slope
Controls the trigger polarity. It is positive to trigger on a rising edge and negative to
trigger on a falling edge.
Key Path:
Trig
Dependencies/
Couplings: Not available for Free Run.
State Saved: Saved in instrument state.
Factory Preset: Positive (rising edge)
Remote Command:
:TRIGger[:SEQuence]:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:SLOPe?
Example: TRIG:SLOP NEG