Chapter 2 163
Instrument Functions: A - L
Input/Output
Instrument Functions: A - L
Remote Command:
[:SENSe]:MIXer:HARMonic:AUTO OFF|ON|0|1 or <boolean>
[:SENSe]:MIXer:HARMonic:AUTO?
[:SENSe]:MIXer:HARMonic <integer>
[:SENSe]:MIXer:HARMonic?
Example: MIX:HARM:AUTO 0
MIX:HARM 8
2.8.9.2 Mixer Type
Selects which type of mixer is in use.
Mixer Type (Presel) activates a tuning signal that is
routed to the
PRESEL TUNE OUTPUT connector on the rear panel of the analyzer. This
signal has a sensitivity of 1.5V/GHz of the LO frequency and drives the tune input of the
HP/Agilent 11974 series of preselected mixers. The sweep rate in
Presel mode is limited to
40 MHz/ms.
Key Path:
Input/Output, External Mixer, Mixer Config
Dependencies/
Couplings: Only available in bands A, Q, U and V, and only when
Mixer Bias is off.
State Saved: Saved in Instrument State
Factory Preset: Unpre
Remote Command:
:INPut:MIXer:TYPE PRESelected|UNPReselect
:INPut:MIXer:TYPE?
Example: INP:MIX:TYPE UNPR