286 Chapter 4
Instrument Functions: P - Z
Trig
Instrument Functions: P - Z
Sweep time
Number of sweep points
The effective trigger offset value will be re-calculated whenever any of these parameters
change.
State Saved: Saved in instrument state.
Factory Preset: 0
500 ms
Range: Hardware specific; dependent upon the ADC being used, current state and
the number of sweep points.
History: Added with firmware revision A.02.00
Remote Command:
:TRIGger[:SEQuence]:OFFSet <time>
:TRIGger[:SEQuence]:OFFSet?
:TRIGger[:SEQuence]:OFFSet:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:OFFSet:STATe?
Remote Command Notes: Trigger offset can only be turned on when in zero span and the
resolution bandwidth is 1 kHz or greater. Trigger offset is available for all
trigger modes.
Example: TRIG:OFFS 100 ms
TRIG:OFFS:STAT ON turns on the trigger offset.