46 Chapter 2
Instrument Functions: A - L
AMPLITUDE / Y Scale
Instrument Functions: A - L
2.1.8.11 dBpT
Sets the amplitude units to dBpT. This is a unit specifically applicable to EMI field
strength measurements. In the absence of a correction factor this unit is treated by the
instrument exactly as though it were dB
µV. You must load an appropriate correction factor
using amplitude corrections for this unit to generate meaningful results. Therefore, this
key is grayed out unless one of the corrections are turned on (in
Amplitude, Corrections
menu) and
Apply Corrections is set to Yes.
Key Path:
AMPLITUDE / Y Scale, More, Y Axis Units
Dependencies/
Couplings: If the grayed out key is pressed an error message is generated.
History: Added with firmware revision A.06.00
Remote Command:
See Y Axis Units on page 41.
Example: UNIT:POW DBPT
2.1.8.12 dBG
Sets the amplitude units to dBG. This is a unit specifically applicable to EMI field strength
measurements. In the absence of a correction factor this unit is treated by the instrument
exactly as though it were dB
µV. You must load an appropriate correction factor using
amplitude corrections for this unit to generate meaningful results. Therefore, this key is
grayed out unless one of the corrections are turned on (in
Amplitude, Corrections menu) and
Apply Corrections is set to Yes.
Key Path:
AMPLITUDE / Y Scale, More, Y Axis Units
Dependencies/
Couplings: If the grayed out key is pressed an error message is generated.
History: Added with firmware revision A.06.00
Remote Command:
See Y Axis Units on page 41.
Example: UNIT:POW DBG