198 Chapter 2
Front-Panel Key Reference
System (Local)
Align Now Accesses the Align Now menu keys and immediately
executes an alignment cycle of the selected system(s).
Menu key selections include:
Key Access:
System, Alignments
All Initiates the alignment of all systems.
Except the Option 1DN Tracking
Generator for models E7402A, E7403A,
E7404A, and E7405A.
Key Access:
System, Alignments, Align
Now
RF
(Ext Cable)
Initiates the alignment of the RF
system with a cable connected from
AMPTD REF OUT to INPUT 50 Ω.
Models E7402A, E7403A, E7404A, and
E7405A.
Key Access:
System, Alignments, Align
Now
RF
Initiates the alignment of the RF
system. This Alignment uses an
Internal Signal Source.
Key Access:
System, Alignments, Align
Now
TG
(Ext Cable)
Initiates the alignment of the tracking
generator only when the Option 1DN is
installedin Agilent E7402A, E7403A,
E7404A, and E7405A. You must
connect a cable from RF OUT 50 Ω to
INPUT 50 Ω
Key Access:
System, Alignments, Align
Now
FM Demod
Initiates an alignment of the FM
Demod portion of the EMC Detectors
and Demod board.
Key Access:
System, Alignments, Align
Now
After a five minute warm-up, Align Now, All (with a
cable connected from AMPTD REF OUT to INPUT 50
Ω) should be performed on Agilent E7402A, E7403A,
E7404A, and E7405A. Refer to your analyzer’s
appropriate Specifications Guide for details on using
Align Now when Auto Align is Off.