EasyManuals Logo

Agilent Technologies E7401A User Manual

Agilent Technologies E7401A
218 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #79 background imageLoading...
Page #79 background image
Chapter 2 79
Front-Panel Key Reference
Det/Demod
Auto The system selects peak detection as the default, but if
a condition arises where a different type of detection
scheme should be better utilized, the system will use
the alternate scheme. For example, the Marker Noise
function uses Average detection when in Auto mode
because the system determines that the data will be
more accurate for noise-type signals.
Key Access:
Det/Demod, Detector Auto Man
Average
Average detection is used when measuring the average
value of the amplitude across each trace interval
(bucket). The averaging method used by the Average
detector will be set to either Video or Power as
appropriate when the Average Type is Auto coupled or
may be set explicitly through the
BW/Avg, Avg Type Auto
Man
key.
Video Average detection computes the averaged trace
interval as appropriate to the Y-axis scale:
Log-power average when in log scale.
Voltage average when in linear scale.
Frequency deviation when Demod View is enabled
for FM Demod.
Power Average (root mean square or RMS) detection
computes the RMS of the samples collected across the
trace interval. The number of samples included in the
RMS average for each trace interval depends on the
sweep time and the resolution bandwidth. Slower
sweep times also increase the number of samples
averaged for each trace interval.
When
Average is selected, Avg appears on the left side
of the display.
Key Access:
Det/Demod, Detector Auto Man
Peak
Enables positive peak detection which displays the
maximum video signal detected during a sampling
interval. Peak detection is selected at power on or by
pressing
Preset when Preset Type is set to Factory and
Power On is set to Preset. When Peak detection is
selected, Peak appears in the upper-left corner of the
screen.
Key Access:
Det/Demod, Detector Auto Man

Other manuals for Agilent Technologies E7401A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E7401A and is the answer not in the manual?

Agilent Technologies E7401A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE7401A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals