42 1200 Series Quarternary Pump User Manual
3 Installing the Pump
Flow Connections of the Quaternary Pump
2 Place the vacuum degasser and the solvent cabinet on top of the quaternary
pump.
3 Put the bottle-head assemblies into solvent reservoirs containing your
mobile phase and place the bottle in the solvent cabinet.
4 Connect the inlet tubes from the bottle-head assemblies to the inlet
connectors A to D (typically the left connection of the channel) of the
vacuum degasser. Fix the tubes in the tube clips of the vacuum degasser.
5 Connect the solvent tubes to the outlet connectors (typically right
connection of the channel) of the vacuum degasser.
6 Connect the syringe adapter from the degasser accessory kit to the solvent
tube of channel A.
7 Using a piece of sanding paper connect the waste tubing to the purge valve
and place it into your waste system.
8 If the pump is not part of a Agilent 1200 Series System stack or placed on
the bottom of a stack, connect the corrugated waste tube to the waste outlet
of the pump leak handling system.
9 Connect the pump outlet capillary (pump to injection device) to the outlet
of the purge valve.

Table of Contents

Other manuals for Agilent Technologies G1314B VWD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1314B VWD and is the answer not in the manual?

Agilent Technologies G1314B VWD Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1314B VWD
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals