88 1200 Series Quarternary Pump User Manual
7 Maintenance
Simple Repairs
Removing the Pump Head Assembly
1 Remove the front cover.
2 Disconnect the Active Inlet Valve cable.
3 Using a 1/4 inch wrench remove the outlet capillary.
4 Disconnect the capillary from the Outlet Ball Valve.
5 Remove the waste tubing and disconnect the Active Inlet Valve tubing.
6 Remove the capillary at the bottom of the Pumphead.
7 Using a 4 mm hexagonal key, stepwise loosen the two Pumphead screws and
remove the Pumphead from the Pump Drive.
When Exchanging the seals
Exchanging the plungers
Exchanging seals of the seal wash option
Tools required Wrench 1/4 inch
4-mm hexagonal key
Preparations Switch off pump at the main power switch and unplug the power cable
CAUTION
Damage of the pump drive
Starting the pump when the pump head is removed may damage the pump drive.
Never start the pump when the pump head is removed.
Ejbe]ZVYhXgZlh

Table of Contents

Other manuals for Agilent Technologies G1314B VWD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1314B VWD and is the answer not in the manual?

Agilent Technologies G1314B VWD Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1314B VWD
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals