54 1200 Series Quarternary Pump User Manual
4 Using the Quaternary Pump
Algae Growth in HPLC Systems
normal wear of the piston seals do not cause the PTFE frits to block over
short-term usage. Please refer to the section “Exchanging the Purge Valve
Frit or the Purge Valve” on page 85 in this manual.
Short lifetime of solvent filters (bottle head assembly). A blocked solvent
filter in the bottle, especially when only partly blocked, is more difficult to
identify and may show up as gradient performance problems, intermittent
pressure fluctuations etc.
Algae growth may also be the possible source for failures of the ball valves
and other components in the flow path.
How to Prevent and/or Reduce the Algae Problem
Always use freshly prepared solvents, especially use demineralized water
which was filtered through about 0.2 µm filters.
Never leave mobile phase in the instrument for several days without flow.
Always discard “old” mobile phase.
Use the amber solvent bottle (part number 9301-1450) supplied with the
instrument for your aqueous mobile phase.
If possible add a few mg/l sodium azide or a few percent organic solvent to
the aqueous mobile phase.

Table of Contents

Other manuals for Agilent Technologies G1314B VWD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1314B VWD and is the answer not in the manual?

Agilent Technologies G1314B VWD Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1314B VWD
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals