80 1200 Series Quarternary Pump User Manual
7 Maintenance
Simple Repairs
7 Position the valve so that the solvent inlet tube connection points towards
the front.
8 Using the 14 mm wrench tighten the nut by turning the valve in its final
position (not more than a quarter turn).
9 Reconnect the adapter at the active inlet valve.
10 Reconnect the solvent inlet tube to the adapter. Reconnect the active inlet
valve cable to the connector in the Z-panel.
11 Reinstall the front cover.
NOTE
After an exchange of the valve it may take several ml of pumping with the solvent used in
the current application, before the flow stabilizes at a %-ripple as low as it used to be when
the system was still working properly.