1200 Series Quarternary Pump User Manual 99
Maintenance
7
Simple Repairs
7
Insert the plungers and carefully press them into the seal. 8 Tighten the lock screw.
9 Slide the pump head assembly onto the metering drive.
Apply a small amount of pump head grease to the
pumphead screws and the balls of the spindle drive.
Tighten the pumphead screws stepwise with increasing
torque.
10 Reconnect all capillaries, tubes and the active inlet valve
cable to its connector.