Agilent ADM Flow Meter Operating Instructions 7
The ADM Flow Meter is powered by three AA alkaline batteries. Power can also be
supplied via the micro USB port located on the side of the unit.
Display
The ADM Flow Meter has a bright OLED display with a battery strength indicator.
The unit automatically adjusts the resolution of the display in response to various
gas flow ranges as follows:
When the battery strength indicator shows that the battery strength is low, replace
the batteries. See “Replacing Batteries" on page 27.
NOTE
Using the USB port to supply power does not recharge the
installed batteries.
NOTE
The ADM Flow Meter can also be powered by rechargeable batter-
ies, but the battery indicator may not display correctly.
Flow (mL/min)
Display resolution
(mL/min)
0 - 9.99 0.01
10.0 - 99.9 0.1
100 - 750 1.0