Installing the GC 1
Agilent Intuvo 9000 GC Installation 37
Figure 4 Remove exterior protective packaging
2 Loosen the captive screws in the shipping door clamp and
remove the clamp. See Figure 4 above.
3 Remove the detector tail brace.
a Use a T20 Torx driver to remove the screws that secure
the MSD frame cover and remove the cover.
b Slide the detector tail brace out of the side of the GC.
c Reinstall the MSD frame cover.
If installing a 7693A tray assembly, you can skip this step
and install the MSD frame cover after installing the tray
mounting hardware.
If installing an MSD or D2 assembly, skip this step.
4 Assemble the Intuvo Torque Driver and attach it to the GC
front door. The calibrated torque driver handle (8710-2790)