SCPI Commands Appendix B
Model N6705 User’s Guide 191
SCPI Command Description
[SOURce:]LIST:TOUTput (continued)
:EOSTep
[:DATA] <Bool> {,<Bool>}, (@chanlist)
Generate triggers at the End Of STep
:POINts? (@chanlist)
Returns the number of EOST list points
:VOLTage
[:LEVel] <NRf> {,<NRf>}, (@chanlist)
Sets the voltage list
:POINts? (@chanlist)
Returns the number of voltage list points
[SOURce:]POWer
:LIMit <NRf+>, (@chanlist)
Sets the power limit on output channels
[SOURce:]RESistance
[:LEVel]
[:IMMediate][:AMPLitude] <NRf+>, (@chanlist)
Sets the output resistance (only on N6781A)
:STATe <Bool>, (@chanlist)
Enables/disables output resistance programming (N6781A)
[SOURce:]STEP
:TOUTput <Bool>, (@chanlist)
Generate a trigger output on the voltage or current step
[SOURce:]VOLTage
[:LEVel]
[:IMMediate][:AMPLitude] <NRf+>, (@chanlist)
Sets the output voltage
:TRIGgered [:AMPLitude] <NRf+>, (@chanlist)
Sets the triggered output voltage
:BWIDth DEFault | FAST1 | FAST2 | FAST3, (@chanlist)
Sets the voltage bandwidth (only on N678xA SMU)
:LIMit
[:POSitive]
[:IMMediate][:AMPLitude] <NRf+>, (@chanlist)
Sets the positive voltage limit (only on N678xA SMU)
:COUPle <Bool>, (@chanlist)
Sets the voltage limit tracking state (only on N6784A)
:NEGative
[:IMMediate][:AMPLitude] <NRf+>, (@chanlist)
Sets the negative voltage limit (only on N6784A)
:MODE FIXed | STEP | LIST | ARB, (@chanlist)
Sets the voltage trigger mode
:PROTection
[:LEVel] <NRf+>, (@chanlist)
Sets the over-voltage protection level
:DELay[:TIME] <NRf+>, (@chanlist)
Sets the over-voltage protection delay (only on N6783A)
:REMote
[:POSitive][:LEVel] <NRf+>, (@chanlist)
Sets the positive remote OV protection level (only on N678xA SMU)
:NEGative[:LEVel] <NRf+>, (@chanlist)
Sets the negative remote OV protection level (only on N6784A)
:RANGe <NRf+>, (@chanlist)
Sets the output voltage range
:SENSe
:SOURce INTernal | EXTernal, (@chanlist)
Sets the remote sense relays
:SLEW
[:IMMediate] <NRf+> | INFinity, (@chanlist)
Sets the output voltage slew rate
MAXimum <Bool>, (@chanlist)
Enables/disables the maximum slew rate override
STATus
:OPERation
[:EVENt]? (@chanlist)
Returns the value of the operation event register
:CONDition? (@chanlist)
Returns the value of the operation condition register
:ENABle <NRf>, (@chanlist)
Enables specific bits in the event register
:NTRansition <NRf>, (@chanlist)
Sets the negative transition filter
:PTRansition <NRf>, (@chanlist)
Sets the positive transition filter
:PRESet
Presets all enable and transition registers to power-on

Table of Contents