Loading...
Using the Source Functions 3
Model N6705 User’s Guide 93
Pulse Parameters
Parameter:
Description:
Start Setting (I
0
or V
0
)
Pulse Setting (I
1
or V
1
)
Delay (T
0
)
Pulse Width (T
1
)
End Time (T
2
)
The setting before and after the pulse.
The amplitude of the pulse.
The delay after the trigger is received but
before the pulse starts.
The width of the pulse.
The time the output remains at the end
setting after the pulse completes.
From the remote interface:
The setting before and after the pulse:
ARB:CURR:PULS:STAR 0,(@1)
ARB:VOLT:PULS:STAR 0,(@1)
The amplitude of the pulse:
ARB:CURR:PULS:TOP 1,(@1)
ARB:VOLT:PULS:TOP 10,(@1)
The delay after the trigger is received but before the pulse starts:
ARB:CURR:PULS:STAR:TIM 0.25,(@1)
ARB:VOLT:PULS:STAR:TIM 0.25,(@1)
The width of the pulse:
ARB:CURR:PULS:TOP:TIM 0.5,(@1)
ARB:VOLT:PULS:TOP:TIM 0.5,(@1)
The time the output remains at the end setting after the pulse:
ARB:CURR:PULS:END:TIM 0.25,(@1)
ARB:VOLT:PULS:END:TIM 0.25,(@1)