Positioning and securing the detector
14
Method 1: Use a sharp knife to cut out the insulation using the scribed
circle as a guide.
Method 2: Pierce the insulation with a screwdriver. Rotate the
screwdriver around the scribed circle to remove excess insulation.
Clean up any pieces of insulation that fall inside the oven.
4. Lower the detector pallet into the cavity. Be sure that the insulation is
in place. Partially tighten the four screws with a T-20 Torx screwdriver.
Retighten all the screws to snugness.
Captive
screws
Insulation