Installing the makeup gas regulator (nonEPC detectors)
11
Installing the makeup gas regulator
(nonEPC detectors)
1. Install a screw in the front or back position of the pneumatics carrier.
Do not tighten the screw.
Install a screw in the makeup gas regulator. Do not tighten the screw.
2. Place the makeup gas regulator in the front or back position of the
pneumatics carrier. Tighten the screws with a T-20 Torx screwdriver.
3. Connect the tube on the makeup gas regulator to the fitting labeled MU
on the flow manifold. Tighten the nut fingertight, then use the 7/16-inch
wrench to tighten it 3/4 of a turn. Follow the same procedure for