Restoring the GC to operating condition
21
9. Reinstall the RFI cover.
10. Replace the pneumatics top cover.
11. Plug in the GC and turn it on.
12. If your GC did not detect the new NPD, check the following connections:
Detector card to main board
Manifold to pneumatics board
Heater-sensor connector
Bead cable to detector card
Signal cable to detector card
If you installed an EPC NPD but you have only temperature control for
the detector, check the connection between the flow manifold and the
pneumatics board.