Chapter 4
Front-Panel Key and SCPI Command Reference
Front-Panel Keys
125
Front-Panel Key and SCPI Command
Reference
4.2 Front-Panel Keys
NOTE Only front panel keys affected by selection of Measuring Receiver mode are described here. For a
complete description of all front panel keys see the PSA Series User’s Guide.
4.2.1 Det/Demod
Accesses all menus that enables to change settings on detectors and demodulation filters.
4.2.1.1 Detector
Sets the type of detectors.
Mode: MRECEIVE
Key Path:
Det/Demod
Remote Command: [:SENSe]:MRECeive:DETector PPEak|NPEak|PNPeak|RMS
[:SENSe]:MRECeive:DETector?
Preset: PPEak
State Saved: Saved in instrument state.
Range: Peak +|Peak −|Peak + -/2|RMS
Dependencies/Couplings: When current measurement is one of the four Audio measurements., the only
available selection is RMS.
Remote Command Notes: Global to the current mode.
Example: MREC:DET PPE
MREC:DET?