EasyManuals Logo

Agilent Technologies PSA Series Manual

Agilent Technologies PSA Series
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #205 background imageLoading...
Page #205 background image
Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
205
Front-Panel Key and SCPI Command
Reference
4.4.4.5.6 Capture Time
Sets the capturing data time. When in manual mode, the user determines the capture time; otherwise, it is
determined automatically by the measurement.
This key is available only when PSA firmware revision A.11.08 or above, or Option 23A is installed.
Key Path: Meas Setup
Remote Command: [:SENSe]:AMDepth:TRIGger:SOURce
IMMediate|IF|EXTernal[1]|EXTernal[2]|RFBurst
[:SENSe]:AMDepth:TRIGger:SOURce?
Default: IMMediate
State Saved: Saved in instrument state.
Range: Free Run (Immediate)|Video (IF Envlp)|RF Burst|Ext Front|Ext Rear
Example: :AMD:TRIG:SOUR IF
:AMD:TRIG:SOUR?
Key Path: Meas Setup
Remote Command: [:SENSe]:AMDepth:CAPTure <time>
[:SENSe]:AMDepth:CAPTure?
[:SENSe]:AMDepth:CAPTure:AUTO 0|1|ON|OFF
[:SENSe]:AMDepth:CAPTure:AUTO?
Preset/Default: 250 ms
State Saved: Saved in instrument state.
Min: 1 μs
Max: 10 s
Resolution 1 ms
Example: :AMD:CAPT 0.25
:AMD:CAPT?
:AMD:CAPT:AUTO OFF
:AMD:CAPT:AUTO?

Table of Contents

Other manuals for Agilent Technologies PSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies PSA Series and is the answer not in the manual?

Agilent Technologies PSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelPSA Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals