Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
217
Front-Panel Key and SCPI Command
Reference
4.4.5.4.2 IF BW Type
Selects the type of using IF BW. It can be automatically set by measurement, or manually set by user. The
function of the two types can reference to FM Deviation’s.
4.4.5.4.3 Average Number
Sets the number that will be averaged. After the specified number (average counts) have been averaged, the
averaging mode (termination control) setting determines the averaging action.
Mode: MRECEIVE
Key Path:
Meas Setup
Remote Command: [:SENSe]:FMDeviation:BANDwidth|BWIDth:TYPE AUTO|MAN
[:SENSe]:FMDeviation:BANDwidth|BWIDth:TYPE?
Preset: AUTO
State Saved: Saved in instrument state.
Range: Auto | Man
Example: :FMD:BAND:TYPE MAN
:FMD:BAND:TYPE?
Mode: MRECEIVE
Key Path:
Meas Setup
Remote Command: [:SENSe]:FMDeviation:AVERage:COUNt <integer>
[:SENSe]:FMDeviation:AVERage:COUNt?
[:SENSe]:FMDeviation:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:FMDeviation:AVERage[:STATe]?
Preset: 25, ON
State Saved: Saved in instrument state.
Min: 1
Max: 8192
Example: :FMD:AVER:COUN 25
:FMD:AVER:COUN?
:FMD:AVER:STAT OFF
:FMD:AVER:STAT?

Table of Contents

Other manuals for Agilent Technologies PSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies PSA Series and is the answer not in the manual?

Agilent Technologies PSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelPSA Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals