Technical Information
J1 – IN-OUT
17
TwisTorr 74 FS AG On-board Controller User Manual / 87-901-054-01 161/192
Signal description
START/STOP: input signal to start or stop the pump. With the
supplied mating connector, the START/STOP (+) signal is connected
to the +24 Vdc pin and the START/STOP (-) signal to the GROUND
pin: in this condition, the pump starts automatically as soon as the
controller recognises the input supply if the interlock is closed ("Plug
& Pump").
INTERLOCK: input signal to control pump rotation. With the
supplied mating connector, the INTERLOCK (+) signal is connected
to the +24 Vdc pin and the INTERLOCK (-)signal to the GROUND pin.
SOFT START: this input is used to "soft start" the pump; in this
condition, the ramp-up time could be up to 20 min.
SPEED SETTING: PWM input signal to set pump speed. PWM signal
characteristics must be as follows:
frequency: 100 Hz +/-20 %
amplitude: 24 V max
duty cycle range: from 25 % to 75 %
corresponding to a pump speed between 1100Hz and 1167Hz
(see the diagram below).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies TwisTorr 74 FS AG and is the answer not in the manual?

Agilent Technologies TwisTorr 74 FS AG Specifications

General IconGeneral
BrandAgilent Technologies
ModelTwisTorr 74 FS AG
CategoryController
LanguageEnglish

Related product manuals