Technical Information
Description
N. Read/ Write Data
Type
Description Permitted Values
204 R N Pump temperature in °C 0 to 110
(60 °C = Threshold
Pump
Overtemperature)
205 R N Pump status Stop = 0
Waiting intlk = 1
Starting = 2
Auto-tuning = 3
Braking = 4
Normal = 5
Fail = 6
206 R N Error code Bit description: see
the following figure
211 R N Controller Heatsink Temperature ( °C) 0 to 110
216 R N Controller Air Temperature ( °C). 0 to 70
224 R A Pressure reading Format = X.X E XX
300 R N Cycle time in minutes (zeroed by the reset
command)
0 to 999999
301 R N Cycle number (zeroed by the reset command) 0 to 9999
302 R N Pump life in hours (zeroed by the reset
command)
0 to 999999
310 to
347
Reserved to Agilent service
400 R A CRC EPROM (QE) QE8XXXX
(where “XXXX” is
variable)
401 R A Boot CRC (BL) BL2XXXX
(where “XXXX” is
variable)
402 R A CRC Param. (PA) PA8XXXX
(where “XXXX” is
variable)
404 R A CRC Parameter structure "XXXX"
184/192 TwisTorr 74 FS AG On-board Controller User Manual / 87-901-054-01