CP-8400 AutoSampler & CP-8410 AutoInjector 11
Setup and Calibration of the CP-8400/8410
WARNING:
BURN HAZARD
The injector nut and locator plate may be very hot during instrument
operation and should not be touched with unprotected hands.
WARNING
Syringe moves automatically during operation of instrument. Keep your
hands and fingers out of the injection area. Failure to do so may result in
accidental needle prick/scratch.
Important Note: In order to accurately calibrate vial positions, a
calibration tool (CP740889) is used. Follow the “vial calibration tool”
procedure given on page 19
Important Note: Using SPME mode sample introduction, it is mandatory
to perform the calibration process with the liquid sampling syringe
provided standard with your Auto Sampler.
Important Note: If one injector is installed, the setup configuration must
reflect the one-injector position to avoid accidentally injecting into a non-
existent injector and causing damage to your GC and bending the
needle.
Injector position
CP-8400/8410 Front Middle Rear
Front position Position 1 Position 2 -
Rear position - Position 1 Position 2