Viscous Sampling Mode
Viscous mode is designed for samples that either are viscous or are
dissolved in a viscous solvent such as cyclohexanol.
Viscous mode actions are summarized below:
The number of pre-injection solvent clean flushes specified in the
method will be performed by drawing a full syringe of solvent from
the specified clean vial and expelling it into the waste cup.
The number of pre-injection sample clean flushes specified in the
method will be performed by drawing a full syringe of sample from
the specified sample vial at 1 µL/sec waiting 9.9 sec with the needle
in the sample, and expelling it into the waste cup at 50 µL/sec.
If an Internal Standard is used, the Internal Standard will be
withdrawn from the specified source vial at this point – the Internal
Standard air gap will be withdrawn if specified.
The volume of sample specified in the Single Method or Method
Sequence Automation will be withdrawn from the specified sample
vial at 1 µL/sec.
The system will pause 9.9 sec, and the sample will be injected into
the specified injector at a rate of 5 µL/sec. The syringe will remain in
the injector for 12 seconds before being withdrawn.
The post injection clean flushes will be executed.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Varian CP-8410 and is the answer not in the manual?

Agilent Technologies Varian CP-8410 Specifications

General IconGeneral
BrandAgilent Technologies
ModelVarian CP-8410
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals