32 03-914852-00:4
Method Sequence Table Parameters
Page 1 (table) * denotes CP-8410
Method Vial Range Inj/Samp Inj Vol 2 Inj Vol 1 Inj Pos
entries 1, 2, ..., 8
0 to 99
(1 to 21)*
1 to 99
0.1 to 5, 10
or 1 to 100
0.1 to 5, 10
or 1 to 100
1, 2, or
Both
default 1
0 to 99
(1 to 21)*
1 1.0 1.0 1
Page 1 of 2
Inj/ Inj Inj Inj
Method Init Final Samp Vol 1 Vol 2 Pos
1 1 20 3 0.5 ---- 1
2 21 38 2 2.0 1.0 Both
3 70 98 1 1.5 ---- 2
Page 2
Use Prep-Ahead
(if Both selected in Setup)
Yes/No
default = No
First Injector Used
Pos. 1, Pos. 2
default = Pos. 1
Use Injection Delay?
Yes or No default = No
Delay Between Injections (min)
0.5-10.0 default = 1
Advance for Second Injection?
Yes or No default = Yes
Clean Between Injections?
Yes or No default = No
Note that the Page 2 Parameters are only used if Both is selected
for the Inj Pos parameter.
The vial range can be set to any combination of vial positions in
ascending order. A range of 1 vial is allowed, thus a range of 1 to 1
means use Vial 1 with the specified method. The default value is 0 to 99
for the CP-8400 and 1 to 21 for the CP-8410.
The number of injections per sample may be set between 1 and 99; the
system will inject the specified number of replicates. If a priority sample is
invoked, the system will remember the number of completed replicates
and return to the next replicate after the priority request is completed.
The default value is 1.