Operation
CP-8400 AutoSampler & CP-8410 AutoInjector 55
Reset Plunger Stroke Counter (Softkey) – Selecting this softkey
causes the Plunger stroke count to be reset to 0.
Injector 1 Needle Depth (%) 0 to 100 default=100
Injector 1 Injection Speed (cm/sec) 1 to 30 default=30
Injector 2 Needle Depth (%) 0 to 100 default=100
Injector 2 Injection Speed (cm/sec) 1 to 30 default=30
The Injector Needle Depth parameters define how far into the designated
injector the needle will penetrate. 100% causes the needle to be
completely inserted to its 2-inch length.
The Injector Injection Speed parameters define the rate at which the
needle will penetrate the injector septum and descend to the set Injector
Needle Depth.
CP-8400/8410 Status
The Automation Status page displays the current state of automation and
the CP-8400/8410 progress through the selected automation table.