15. Output modules
70
The following trigger functions can be programmed as variants :
0: Start / stop a measuring operation 0: Start - Stop
1: Once-only manual measuring point scan 1: Once-only scan
2: Clear all maximum values and min values 2: Clear max and min values
3: Print measured value 3: Print
4: Start / stop a measuring operation on level-controlled basis
4: Start - Stop, Level-con-
trolled
8: Zero-set a measured value 8: Zero-set meas. value
-5: Execute macro 5 (see manual 6.6.5) -5: Macro 5
-6: Execute macro 6 -6: Macro 6
-7: Execute macro 7 -7: Macro 7
-8: Execute macro 8 -8: Macro 8
-9: Execute macro 9 -9: Macro 9
15.3 Analog outputs
V5 output modules
For the purposes of analog recording of measured
values it is still possible, at sockets A1 and / or A2 (2),
to connect V5 output modules with an analog output
controlled by the device, e.g. recording cable
ZA1601-RK (see Manual 5.1.1).
To select the socket press : <P> : ▲ or ▼
The following output modes can be programmed as variants :
This gives, depending on the analog output, the following output signals :
Voltage output -1.2 ... +2.00 V 0.1 mV / digit
Voltage output -6.0 ... +10.0 V 0.5 mV / digit
Current output 0.0 ... 20.0 mA 1 µA / digit
In variant 2 ´Assigned internally´, after selecting the Mxx function,
it is possible to program the measuring point to be output :
2: Assigned internally
M
V6 output modules
The new V6 relay trigger analog adapter ZA8006-
RTA3 (see Manual 5.1.3) offers the option at ports
4 to 7 of up to 4 in the output signal separately
configurable external analog outputs.